How To Calibrate A Masterflex L/S Digital Drive

Peristaltic Pump Calibration

A Masterflex L/S digital drive's speed setting tells the pump how fast to turn, not how much liquid it's actually moving. The real flow rate depends on the exact tubing you have loaded, and tubing varies enough from batch to batch that the number on the display can drift from the number in your beaker. Tubing calibration closes that gap by measuring what the pump actually delivers and adjusting the drive to match, and it takes only a few minutes with a beaker of water and a graduated cylinder.

At A Glance

Best suited for: any Masterflex L/S digital drive paired with L/S pump tubing

What you need: a beaker or container of water, a graduated cylinder, the pump head and tubing you'll actually use

Time required: a few minutes, plus one or two repeat runs until the reading matches

Main benefit: turns a raw speed setting into a verified, real world flow rate

What You'll Need

You'll need a beaker or container of water, a graduated cylinder, a Masterflex pump head and Masterflex tubing. The demonstration below uses LS15 silicone tubing. A clamp and stand aren't strictly necessary, but they make the setup steadier and easier to work with. Dyeing the water makes the level easier to read against the light background of a graduated cylinder, any safe method works, it's purely a visibility aid.

Calibration setup with a beaker of red dyed water, Masterflex pump head, tubing and a graduated cylinder on a lab bench

A typical calibration setup: beaker, pump head with tubing, and a graduated cylinder to catch the output.

Turn On The Drive And Open Setup

Turn on the drive, then go to the main menu by selecting the setup icon and pressing Enter. Use the up and down arrow keys to move through the menu until Tubing Cal is highlighted, then press Enter to select it.

Hand reaching to turn on the Masterflex L/S digital drive with the pump already connected to tubing

Turning on the drive before entering the setup menu.

Masterflex L/S digital drive display showing the setup menu with Flow Units, Tubing Size, Flow Rate, Tubing Cal and Pump Direction options

The setup menu, use the arrow keys to highlight Tubing Cal.

Set Your Flow Direction, Tube Size And Flow Rate

Inside the Tubing Cal screen, set the drive for your desired flow direction, tube size and flow rate. The calibration volume itself is fixed at this stage and can't be changed here.

Masterflex L/S digital drive Tubing Cal screen showing a 180 mL calibration target and 500 mL per minute flow rate setting

The Tubing Cal screen, set flow rate and tube size here before starting.

Calibrate at your real flow rate: it's best to calibrate at the same flow rate you'll actually be using in your application. That's what gives you the most accurate result for the work the pump is actually going to do, rather than a generic calibration at an arbitrary speed.

Prime The Pump

Press and hold the prime key to prime the pump and fill the tubing with water. Once the tubing is primed, empty any water that's collected in the graduated cylinder back into the beaker, this ensures an accurate measurement once the actual calibration run starts.

Hand pressing the prime key on the Masterflex L/S digital drive while water moves through the tubing

Run The Calibration And Read The Result

Highlight the Start field and press Enter to begin the calibration run. After the pump stops automatically, read the amount collected in the graduated cylinder, being sure to read from the bottom of the meniscus for an accurate result. In the video's own example, the first run delivered 194 mL.

Masterflex L/S digital drive running a calibration cycle, Tubing Cal screen showing a 180 mL target at 500 mL per minute, with red dyed water moving through the tubing into a graduated cylinder

The calibration run in progress.

Graduated cylinder showing a red liquid level being read against the millilitre markings after a calibration run

Reading the collected volume from the bottom of the meniscus.

Adjust The Cal Volume, Then Verify

Once you've read the cylinder, the Cal Volume field on the display will be highlighted. Press Enter, adjust the value to match your actual reading, then press Enter again to confirm it. Empty the cylinder back into the beaker and run the calibration again to check the pump has adjusted correctly.

This can take more than one pass. In the video's example, the first run read 194 mL, the second run after adjusting read 181 mL, and it took a third run to land on the target of 180 mL and confirm the calibration as successful. If your result is still inaccurate after a run, repeat the sequence, don't stop at the first adjustment.

Masterflex L/S digital drive Tubing Cal screen with a digit in the calibration volume field highlighted for editing Graduated cylinder showing the liquid level from a verification calibration run after the volume was adjusted

A repeat run confirms whether the adjustment landed on target.

Lock In The Calibration

Once a run comes back accurate, highlight Exit and press Enter to lock in the calibration. Scrolling down to Exit and pressing Enter again takes you back to the mode you were previously in. At that point, a small "C" appears in the bottom left corner of the display, confirming the pump is now calibrated.

Masterflex L/S digital drive display back in continuous mode after calibration, showing flow rate in millilitres per minute

If your calibration volumes vary significantly between runs: follow the break-in procedure for new tubing first, running it at full speed for 10 minutes before calibrating. New tubing can behave inconsistently until it's been run in.

Need Help Setting Up A Masterflex Pumping System?

Choosing the right drive, pump head and tubing depends on your flow rate, chemical compatibility and duty cycle. John Morris Group can help configure a Masterflex system for your application.

Contact John Morris Group

Or call 1300 501 555

Masterflex L/S Digital Drive, 0.1 to 600 rpm, 115/230 VAC

Masterflex L/S Digital Drive, 0.1 to 600 rpm (Avantor Fluid Handling)

View This Drive
September 1, 2026

Learn more about working at John Morris