The Minitron offers many advantages for both bacterial and cell cultures such as a quiet but powerful drive placed outside of the incubation unit, easy cleaning and a compact design.
Technical data
- Dimensions (L x D x H) 800 x 625 x 700 mm (one unit on rubber feet)
- Stackable (up to 2 units)
- Available shaking throw 25 mm or 50 mm
- Rotation speed range 20 - 400 min-1
- Max. load 12 kg
- Standard parameters: Temperature and rotation speed
- Optional parameters: Cooling, CO2, humidity
- Internal illumination: Available
- Data transmission: Ethernet connection to the bioprocess platform software eve®
Drive with 25 mm shaking throw
Rotation speed (max. values)
The actual rotation speed that can be achieved depending on different factors, such as loading,
vessel types or holder. The following max. speeds are based on an optimum load given when using
12 x 500 mL Erlenmeyer flasks without baffles.
Humidification: Hygienic Humidity Control (HHC)
The Hygienic humidity control (HHC) allows to recreate in the incubation chamber the appropriate growing conditions for cells up to 75 % rH and strongly prevents evaporation from shake flasks, deep well plates and microtiter plates.
CO2 control (0-20 %)
Via this option the CO2 concentration is electronically regulated in the incubation chamber allowing a very precise control for an optimal cultivation of mammalian cells or algae.